510(k) Data Aggregation

    K Number
    K970333

    Validate with FDA (Live)

    Date Cleared
    1998-01-28

    (365 days)

    Product Code
    Regulation Number
    866.2560
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BACTEC® MYCO/F Lytic culture vials when used with the 9000 Blood Culture series of instrumentation are intended as an adjunct to routine blood culture for patients suspected of having mycobacteria, yeast and fungi septicemia. Extended incubation times (7 days for yeast, 30 days for fungi, and 42 days for mycobacteria) will permit recovery of mycobacteria and fungi when more rapidly growing organisms are not present. This medium may also be used for the culture of sterile body fluids when yeast or fungi are suspected.

    Device Description

    BACTEC MYCO/F LYTIC Culture medium is a Middlebrook 7H9 and Brain Heart Infusion broth formulation for the recovery of mycobacteria from blood specimens and fungi from blood and sterile body fluids. The range of specimen volume which can be cultured is one to five mL. with optimum recovery obtained at three to five mL. Specific modifications were made to enhance the growth and recovery of mycobacteria, veast, and fungi. These modifications include ferric ammonium citrate to provide an iron source for specific strains of mycobacteria and fungi, the addition of saponin as a blood lysing agent and the addition of specific proteins and sugars to provide nutritional supplements. Each vial contains a sensor which can detect decreases in oxygen concentration in the vial resulting from microorganism and growth. The sensor is monitored by the BACTEC 9000 Blood Culture Systems for increasing fluorescence which is proportional to the decrease in oxygen. A positive determination indicates the presumptive presence of viable microorganisms in the vial.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    The BACTEC MYCO/F LYTIC Blood Culture Medium is deemed substantially equivalent to predicate devices (BACTEC 13A for mycobacteria and BACTEC NR FUNGAL for yeast/fungi) based on internal and clinical studies demonstrating "equivalent performance." While explicit numerical acceptance criteria for recovery rates or time to detection are not provided as fixed thresholds, the studies aim to show that the new device performs comparably or better than the predicate devices across various clinically relevant scenarios.

    Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implied by Substantial Equivalence Goal)Reported Device Performance (BACTEC MYCO/F LYTIC)Predicate Device Performance (BACTEC 13A / BACTEC NR FUNGAL)Notes
    Recovery of Mycobacteria (Clinical)Equivalent or improved recovery compared to BACTEC 13A.10 (9%) isolates recovered only by Myco/F Lytic.3 (3%) isolates recovered only by BACTEC 13A.Myco/F Lytic demonstrated improved unique recovery for mycobacteria. Total 111 pathogenic mycobacteria isolates.
    Recovery of Yeast & Fungi (Clinical)Equivalent or improved recovery compared to ISOLATOR™ System.7 (22%) isolates recovered only by Myco/F Lytic.6 (19%) isolates recovered only by ISOLATOR™ System.Myco/F Lytic demonstrated slightly improved unique recovery for yeast/fungi. Total 32 pathogenic yeast/fungal isolates.
    Recovery of Yeast & Fungi (Internal)Equivalent recovery and time to detection (TTD) compared to BACTEC NR FUNGAL.Improved recovery of Histoplasma capsulatum and Malesezzia furfur. Equivalent for various Candida and Cryptococcus species. Not detectable for Penicillium purpurescens and Blastomyces dermatitidis. Inconsistent for Hansenula anomala, Exophiala jeanseimelei, Actinomyces bovis, Rhodotorula rubra, Mucor ramosissimus at low inoculum.(See Table 2 for detailed TTD)Mixed results depending on the specific organism and inoculum level. The phrase "equivalent performance" might be an overarching claim, with noted exceptions.
    Recovery of Mycobacteria (Internal)Acceptable recovery and TTD compared to BACTEC 13A (implicitly).Acceptable for a majority of tested species. Detection delays/compromised recovery for M. intracellulare, M. malmoense, M. haemophilum, M. xenopi with < 3 mL blood.(See Table 3 for detailed TTD)Generally acceptable, with specific limitations noted for low blood volumes.
    False Positive Rate (Clinical)Acceptable rate for a blood culture system (no explicit number given).0.7% (11 out of 1,488 blood cultures)Not directly compared to a predicate here.Within expected range for such systems at the time.
    False Negative Rate (Clinical)Acceptable rate for a blood culture system (no explicit number given).0.07% (1 out of 1,488 blood cultures)Not directly compared to a predicate here.Within expected range for such systems at the time.
    Contamination Rate (Clinical)Acceptable rate for a blood culture system (no explicit number given).3.3%Not directly compared to a predicate here.Within expected range for such systems at the time.

    Study Details

    2. Sample sizes used for the test set and the data provenance:

    • Clinical Studies (Test Set):
      • Mycobacteria: 1,100 blood culture specimens.
      • Yeast and Fungi (Blood): 748 blood culture specimens.
      • Overall Clinical Evaluation (Mycobacteria, Yeast, Fungi, Other Bacteria): 1,488 blood cultures.
    • Internal Performance Studies (Test Set - In Vitro):
      • Yeast and Fungi: A variety of species at different CFU levels (e.g., TNTC, 170, 16, 80, 6 CFU/vial) and specimen volumes (0 mL, 1 mL, 5 mL blood). Each value is the average of "all vials per test condition" (specific number of vials per condition not detailed).
      • Mycobacteria: A variety of species at different CFU levels (e.g., 0, 1, 2, 3, 5, 7, 12, 13, 16, 20, 31, 44, 45, 49, 53, 58, 68 CFU/vial) and specimen volumes (1 mL, 3 mL, 5 mL blood). Two replicates ("Replicate" and original reading) are shown for each condition.
    • Data Provenance:
      • Clinical Studies: Conducted at "two clinical sites considered large tertiary care teaching hospitals in geographically diverse areas" (for mycobacteria) and "four clinical sites considered large tertiary care teaching hospitals" (for fungi). Patient populations included HIV, immunocompromised, transplant patients, and those suspected of mycobacterial or fungal infections. This indicates prospective clinical data from diverse, real-world patient samples.
      • Internal Performance Studies: Conducted by Becton Dickinson Microbiology Systems. These are retrospective/in-vitro laboratory studies using controlled inoculums of specific organism strains.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The document does not explicitly state the number of experts or their specific qualifications (e.g., radiologists with X years of experience). However, for the clinical studies, ground truth was established by:

    • Determining the "total number of pathogenic mycobacteria isolates recovered" and "total number of pathogenic yeast and fungal isolates recovered" from the various culture media.
    • For false positive/negative rates, instrument results were compared to "smear and/or subculture-negative" or "smear and/or subculture-positive." This implies standard microbiological techniques and interpretation by laboratory professionals, rather than image interpretation by radiologists.

    4. Adjudication method for the test set:

    The document does not describe a formal adjudication method (like 2+1 or 3+1). The ground truth for the clinical studies appears to be based on the established clinical laboratory results from the two or four tertiary care hospitals involved, likely following standard diagnostic protocols of the time. For false positives/negatives, smear and/or subculture results served as the determinant.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable to this device. The BACTEC MYCO/F LYTIC system is a blood culture medium used with an automated instrument for microbial growth detection, not an AI-assisted diagnostic imaging device that involves "human readers" in the traditional sense of image interpretation. The comparison is between different culture media/systems and their ability to recover and detect microorganisms.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The BACTEC MYCO/F LYTIC is a culture medium used with an instrument (BACTEC 9000 Blood Culture Series). The instrument automates the detection of growth (based on oxygen metabolism), but the "algorithm only" concept usually refers to AI in image analysis or similar fields. The performance measured is that of the medium and instrument system in detecting growth. While automated, human intervention is still required for specimen collection, inoculation, and interpretation of positive results (e.g., Gram stain, subculture for identification). Therefore, it's a device-only performance rather than a standalone algorithm in the AI sense.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Clinical Studies: Ground truth was established by microbiological culture results from the comparative methods (BACTEC 13A, ISOLATOR™ System) and confirmed by subculture and/or smear for positive/negative determinations. For identifying pathogenic isolates, standard laboratory identification methods would have been used.
    • Internal Performance Studies: Ground truth was established by known inoculum levels of specific microbial strains (CFU/vial) in a controlled laboratory setting.

    8. The sample size for the training set:

    • The document does not explicitly describe a "training set" in the context of machine learning. The studies are traditional clinical and internal performance evaluations demonstrating the device's efficacy and substantial equivalence.
    • The "Substantial Equivalence" section mentions "Internal and clinical studies demonstrated equivalent performance with all predicate devices," implying these studies were used to demonstrate performance rather than to "train" an algorithm.

    9. How the ground truth for the training set was established:

    • As there's no explicitly defined "training set" in the AI/ML context, this question is not applicable in the sense of how an algorithm's ground truth was established by experts for training.
    • The "ground truth" for the overall evaluation involves the known presence/absence of microorganisms and their identification through standard microbiological laboratory procedures and controlled laboratory inoculations, as described in point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K970512

    Validate with FDA (Live)

    Date Cleared
    1998-01-08

    (331 days)

    Product Code
    Regulation Number
    866.2560
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BACTEC ® MYCO/F LYTIC Culture vial when used with the BACTEC 9000MB System is a qualitative test for the culture and recovery of mycobacteria in human blood speciments from patients who are suspected of having septicemia.

    Device Description

    BACTEC MYCO/F LYTIC blood culture medium is a non-selective growth medium intended for the culture and recovery of mycobacteria and designed for blood volumes of one to five mL. BACTEC MYCO/F LYTIC culture medium is a Middlebrook 7H9 and Brain Heart Infiusion broth formulation with specific formulation modifications made to enhance the growth of mycobacteria. It is used specifically with the BACTEC 9000MB instrument in the monitoring of clinical blood specimens for the presence of microorganisms. This medium contains the same fluorescence senor as the BACTEC MYCO/F Sputa culture medium and detection is based on changes in oxygen concentration in the vial resulting from metabolism and growth of microorganisms. The sensor is monitored by the BACTEC 9000MB System for increasing fluorescence which is proportinal to the decrease in oxygen. A positive determination indicates the presumptive presence of viable microorganisms in the vial.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the BACTEC MYCO/F LYTIC Culture Medium, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly define formal "acceptance criteria" with specific thresholds (e.g., "sensitivity must be >X%"). However, it implies acceptance criteria by comparing the performance of the BACTEC MYCO/F LYTIC medium to a predicate device (BACTEC 13A) and by reporting internal performance characteristics.

    The implied objective is that the BACTEC MYCO/F LYTIC medium should be at least comparable to, or ideally better than, the predicate device in terms of mycobacteria recovery and detection, with acceptable false positive and false negative rates.

    Since explicit acceptance criteria are missing as numerical targets, I will present the key performance metrics reported in the studies.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance (BACTEC MYCO/F LYTIC)
    Clinical Recovery (Overall)Comparable or improved recovery of pathogenic mycobacteria compared to BACTEC 13A.Total 39 pathogenic mycobacterial isolates recovered. 5 (13%) recovered ONLY in BACTEC MYCO/F LYTIC. 2 (5%) recovered ONLY by BACTEC 13A. 32 recovered by both.
    Clinical Recovery (Species-specific)Demonstrate recovery of various pathogenic mycobacteria species.Mycobacterium avium: 3/30 recovered only by MYCO/F LYTIC, 1/30 only by 13A, 26 by both. Mycobacterium tuberculosis: 6/6 recovered by both. Mycobacterium kansasii: 2/3 recovered only by MYCO/F LYTIC, 1/3 only by 13A.
    Internal Recovery (Species Diversity)Detect a variety of mycobacteria species as positive.Detected: M. tuberculosis, M. kansasii, M. fortuitum, M. intracellulare, M. bovis, M. terrae, M. simiae, M gordonae, M. celatum, M. abscessus, M. maimoense. Unsatisfactory recovery: M. xenopi and M. szulgai.
    False Positive RateAcceptably low false positive rate. (No explicit numerical threshold given, but generally, lower is better).0.4% (1 out of 284 blood specimens). For instrument positive MYCO/F LYTIC vials: 2.6% (1 out of 38). (Note: 16 false positives out of 28 overfilled vials were excluded from the study due to protocol deviation).
    False Negative RateAcceptably low false negative rate. (No explicit numerical threshold given).0% (Based on terminal subcultures of 50% of negative vials).
    Contamination RateAcceptably low contamination rate. (No explicit numerical threshold given).0.9%
    Time to Detection (Internal Study)General indication of detection within a reasonable timeframe (implied, not explicitly quantified as a criterion for clinical relevance).Varies by species and blood volume. Examples: M. tuberculosis avg 22.9 days (1 mL), 18.7 (3 mL), 17.3 (5 mL). M. avium avg 8.0-8.1 days. M. fortuitum avg 8.0-5.1 days. (See Table 2 for full details).

    2. Sample Size and Data Provenance

    • Clinical Test Set Sample Size: 284 blood specimens.
    • Data Provenance: Prospective clinical study conducted at "two clinical sites considered large tertiary care teaching hospitals in geographically diverse areas." The specific country of origin is not mentioned but typically for FDA submissions of this era, the studies would be conducted in the United States.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not explicitly state the number of experts or their specific qualifications for establishing the "ground truth" (e.g., "smear and/or subculture-negative/positive") for the clinical test set. It mentions "microbiological workup" and validation against "smear and/or subculture." This implies standard laboratory practices performed by qualified laboratory personnel, but details on expertise (e.g., years of experience, specific certifications) are absent.

    4. Adjudication Method

    The document does not describe an explicit adjudication method (like 2+1 or 3+1) for resolving discrepancies in the ground truth for the clinical test set. The ground truth ("smear and/or subculture-negative/positive") appears to be determined by confirmed laboratory results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This study compares a device (the BACTEC MYCO/F LYTIC medium) directly against a predicate device (BACTEC 13A medium) in a clinical evaluation, but it is not a "human reader" study; it's a comparison of culture media performance. Therefore, an effect size of human readers improving with or without AI assistance is not applicable.

    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone performance study was conducted. The "INTERNAL PERFORMANCE" section and "Table 2: Detection of Mycobacteria in the Myco/F Lytic Medium" represent a standalone evaluation. This study assessed the recovery and time to detection of various mycobacteria species at different CFU levels specifically with the BACTEC MYCO/F LYTIC medium and the BACTEC 9000MB instrument, without direct comparison to human reading or other media as the primary objective.

    7. Type of Ground Truth Used

    • Clinical Study: The ground truth for the clinical performance evaluation was established through laboratory confirmation (smear and/or subculture results). This represents definitive microbiological results, which would fall under a form of "pathology" in a broader sense of laboratory diagnostics.
    • Internal Study: The ground truth for the internal performance study (Table 2) was based on known mycobacteria strains and quantified CFU/bottle (Colony Forming Units per bottle), with the outcome being the time to detection by the BACTEC 9000MB instrument.

    8. Sample Size for the Training Set

    The document does not mention a training set for the BACTEC MYCO/F LYTIC culture medium. This type of device (a culture medium for microbial growth detection) does not typically involve machine learning algorithms that require explicit "training sets" in the conventional sense. The "internal performance" study (Table 2) could be considered an initial validation or characterization of the medium's performance with known isolates under controlled conditions, serving a similar purpose to testing the device's inherent capabilities before clinical evaluation.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no explicit mention of a "training set" in the context of machine learning, this question is not directly applicable. If the "internal performance" study is considered a foundational or characterization study, the ground truth was established by using known, cultured mycobacteria strains with quantified concentrations (CFU/bottle).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1